ELREC T
Versatile IP or Resistivity System
Features
General
The Elrec T is truly a crossover instrument, bridging the
capabilities of both a resistivity meter and an IP receiver. As a resistivity meter it
offers all of the functions and capabilities as described for the
Syscal
R2. As an IP receiver it offers the functions and specifications of the R2,
as well, but as a two dipole receiver.
As a resistivity meter, functioning in the
transmitter/receiver mode, the Elrec T may be powered with either a 125, 250, or 1200 Watt
converter. For a complete description of these capabilities please refer to the Syscal R2
specifications. The Elrec T offers the same receiver specifications for resolution,
stacking, and memory operation. It is also compatible with the Intelligent Node automated
switching system.
For deeper investigations the Elrec T may be used as a
conventional IP receiver, offering two dipoles. In the receiver only mode it may be used
with any time domain transmitter such as the VIP 3000, VIP 4000, or VIP 10000. In this
mode the Elrec T offers the same specifications as the Elrec 2. Synchronization is
automatic with a 10 ms sampling rate.
Specifications
-
Maximum Output Voltage: 800V (1600 V peak-to-peak)
-
Maximum Output Current: 2.4A
-
Output Current Specifications:
-
Resolution: 10µA Accuracy: standard 0.3% (max 1% from -20°C
to +70°C)
-
Output Current Waveform:
-
Frequency domain: ON+, ON- for resistivity Time domain: ON+,
OFF, ON-, OFF for resistivity and chargeability
-
Pulse Duration: (ON time) programmable from 0.25 to 10s, with
preset values of 0.5s, 1s and 2s
-
Overheating Protecting: Thermal circuit breaker
-
DC Power Source
-
The external DC source for ground energization can be:
-
125W DC/DC converter (400V, 4kg) with battery
-
250W DC/DC converter (800V, 5kg) with battery
-
1200W AC/DC converter (800V, 25 kg) with 220V (or 110V) motor
generator
Receiver
-
Input Impedance: 10Mohms
-
Input over voltage Protection
-
Input Voltage Range: -5V to +5V
-
Automatic SP Bucking: -5V to +5V with linear drift correction
(up to 1 mV/s)
-
Power Line Rejection: 50 Hz and 60Hz
-
Ground Resistance Measurement: from 0.1 to 1,000 Kohms
Voltage Measurement:
Resolution: 1µV after stacking Accuracy: standard 0.3% (max
1% from -20°C to +70°C)
Chargeability Measurement:
Resolution: 0.1 mV/V Accuracy: 1% of displayed value for a
voltage greater than 10mV
Stacking: Continuous digital stacking up to 250 stacks
Two Dipole Receiver Mode
-
Requires a Time Domain waveform transmitter with 0.5, 1 or 2s
pulse duration
-
Automatic synchronization with 10ms sampling rate
-
Common mode rejection between channels 86dB (for RS=0)
Standard Components
The ELREC T console, data transfer cable, instruction manual
and shipping case.
Ordering Information
|
Description
|
|
Order Number
|
| ELREC T |
|
500-190-0025 |
|
250W Converter |
|
500-190-0009 |
|
 |
 |